LS1 LSR1

Technical Specification

Power
LS1.1
LS1.2
Mains voltage
230 Vac, 50 Hz
120 Vac, 60 Hz
230 Vac, 50 Hz
Power consumption
4 A (max)
8 A (max)
4 A (max)
Over-voltage
Category II,
2,500 V
Category II,
2,500 V
Category II,
2,500 V
Class/type
1
1
1
Earth leakage current
<500uA
<500uA
<500uA
Fuses
Main fuse
2xT4 AH 250 V
2xT10 AH 250 V
2xT8 AH 250 V
Compressor fuse

T4 AH 250 V

T8 AH 250 V
T8 AH 250 V
Heater fuse

T4 AH 250 V

T8 AH 250 V
T8 AH 250 V
Classification
Safety requirements according to
SS-EN 61010-1
Electromagnetic compatibility
61326-1:2006
Conducted Emission: CISPR 11 Class A, voluntary class B.
Radiated Emission: CISPR 11 Class A
Encapsulation class
IP21
Dimensions and weight
Width
680 mm
Depth
764 mm
Height (main unit/lowest operating position)
1,116 mm/1,755 mm
Weight (without disposable set)
135kg
Other
Specification
Accuracy
Pump
0-6 liter/min
+- 5 %, minimum 0,05 l/min
Temperature measurement
0-45C°
+- 0,3 °C
Pressure measurement, artery and veins
– 10 mmHg to 50 mmHg
+- 2 mmHg
Pressure measurement, system pressure
– 10 mmHg to 300 mmHg
+- 2 mmHg
Noise level, alarm
68dB(A)
-
Maximum noise level
68dB(A)
-
Maximum connected gas pressure permitted - O2
7 bar
-
Maximum connected gas pressure permitted - N2/CO2
7 bar
-
Gas flow O2
<3 l/min
-
Gas Flow  N₂/CO₂ 
<15 l/min
-
Indoor environment
Ambient temperature during operation
min/max:  15/ 40 C°
Relative air humidity during operation (non-condensing)
max 85%